UVP 03

UV-C irradiance probe (with gain adjustment) and temperature monitoring. Made of AISI 316 stainless steel, equipped with ¼ ” male thread (for direct mounting of the tube on its female nut) and silicone O-rings for a perfect hydraulic seal.
UV-C irradiance at a wavelength of 253.5 nm is “captured” by a selective photodiode inside the casing, isolated from water by a quartz cylinder that allows perfect irradiance readout. Gain adjustment of the read UV-C signal allows this probe to be used with all existing tubes regardless of the number of lamps inside. Temperature is measured through a sensor inside the casing, which outputs a proportional voltage signal.
UVP 03 has a 5 V dc power supply and connects to the logic with a standard 4,5-meter shielded cable with a straight connector and shield mounted on it for interference elimination and is available in different enclosure.